Model Number |
B-2000 |
SPECIFICATIONS: Model: B-2000 POWER Power Requirements: AC220V 10%, 50Hz 2% Power Input: ≤450VA Refer to IEC60601-1, this equipment belongs to Class I, Type B. Its operation is Continuous. ALARM Power Alarm Power Failure: Activates if primary power to the Incubator Fails or the power wire is accidentally Disconnected from the wall receptacle.
Fan Alarm Fan Failure: Activates if Fan motor malfunctions. Sensor Alarm Air Sensor Failure: Air Sensor malfunction, refer unit to service. Sensor Out of Position: Activated when Sensor is not in correct position To perform calibration or continue operation. Skin Sensor Disconnected: Activated during Skin Control Mode when Skin Sensor is disconnected from Temperature Controller. Skin Sensor Failure: Activated when Skin Sensor malfunctions. Deviation Alarm High Air Temperature: Activated if displayed Air Temperature is 1.5 centigrade above set point Low Air Temperature: Activated if displayed Air Temperature 2.5centigrade below set point
Skin Temperature: Activated if displayed Skin Temperature is 1.0 centigrade above set point Low Skin Temperature: Activated if displayed Skin Temperature is 1.0Over Temperature Alarm Over Air Temperature: Activated when displayed Air TemperatureWas approach 38centigrade for set temperature ≤37centigrade, Or approach 40centigrade for set temperature >37centigrade in Air Mode.
Over Skin Temperature: Activated when displayed Skin Temperature was approach 38centigrade for set temperature ≤37centigrade, Or approach 39centigrade for set temperature >37centigrade in Skin Mode. System Alarm System Failure: Activated if Temperature Control System malfunctions.
TEMPERATURE CONTROLRANGEAND CORRELATIVE SPECIFICATIONS Air Temperature: 20centigrade ~37centigrade, for set temperature ≤37centigrade 37centigrade ~39centigrade, for set temperature >37centigrade Skin Temperature: 20centigrade ~37centigrade, for set temperature ≤37centigrade 37centigrade~38centigrade, for set temperature >37centigrade Temperature Rise Time: ≤30min Temperature Variability: 0.5centigrade Temperature Uniformity: ≤ 1 centigrade Skin Sensor Precision: ≤0.3centigrade ENVIRONMENT REQUIREMENT
Environment Temperature Operation Temperature Range+20centigrade ~+30centigrade Store Temperature Range -40centigrade ~+70centigrade Environment Humidity Operation Humidity Range: 30%~75% Store Humidity Range: ≤80% Atmospheric Pressure Range: 500hPa~1060hPa OTHER SPECIFICATION Noise Level Within Hood Environment ≤55dB(Environment noise below 45dB Carbon Dioxide (CO2) Level within the Hood Less than 0.5% when a mixture of CO2
In air is delivered at 750ml/min At point 10cm above the center of the mattress. Air Velocit y Over Mattress less than: 0.35m/s Infant Bed Max Load: 10kg Transfusion Shelf Max Load: 2kg Shelf Max Load: Cabinet Max Load: 6kg
|