Portable Ventilator(JIXI-H-100C)
Scope Use for first aid Use for ambulance Use for transport
Physical Specifications Dimensions Carrying packages Dimensions Packaging Dimensions Width: 300mm Width: 460mm Width: 620mm Depth: 160mm Depth: 200mm Depth: 240mm Height: 230mm Height: 400mm Height: 410mm Weight 3Kg Weight 7Kg Weight 11Kg
Environmental Specifications Operating Requirements Storage/Shipping Requirements Temperature: 5 to 40oC Temperature: -20 to 70oC Relative Humidity: less than 80% Relative Humidity: less than 93% The atmosphere pressure: 70 to 106 kPa The atmosphere pressure: 70 to 106 kPa
Electrical Specifications Input Voltage Requirements AC Voltage: 110V~240V DC Voltage: 12V Inlet Battery:12V , 2Ah (Can be used for 8 hours) Frequency: 50Hz/60Hz Power: 2A
Pneumatic Specifications Pneumatic Specifications Oxygen Inlet Pressurized Oxygen: 0.4MPa medical oxygen or use 2L oxygen cylinder (Choice of suppliers)
Performance Specification And Tolerances Controls Power ON/OFF Modes CMV(IPPV) A/C(SIPPV) SIMV SPONT PEEP(Choice of suppliers) Tidal Volume 50~1200ml Breath Rate 5~60 bpm PEEP 0~2KPa O2% 45~90% Sigh 1~10 per 100 breath Peep 10-60 hpa (Choice of suppliers) Trigger Level -10~0hpa Over Pressure Relief <6kpa
Alarms and Indicators High Pressure Alarm Limit 2~6kpa flash, red indicator Low Pressure Alarm Limit 0. 2~2kpa Red indicator Alarm Silence 120 sec Monitors
Tidal Volume 50~1200ml Total Breath Rate 5~60 bpm Main Circuit Pressure -1~6KPa Wave Display Main Circuit Pressure What Is Provided
Part Name Number Specification Ventilator 1 Oxygen Cylinder 1 2.5L can be used for 1 hour Breath Circuit 1 1.2m Carrying Package 1 all parts in it Breath Value
|