Scope
Use for first aid Use for ambulance Use for transport

Physical Specifications
 Dimensions Carrying packages Dimensions Packaging Dimensions
 Width:    300mm Width:    460mm Width:    620mm
 Depth:    160mm Depth:    200mm Depth:    240mm
Height:    230mm  Height:    400mm  Height:    410mm
Weight    3Kg Weight    7Kg Weight    11Kg

Environmental Specifications
Operating Requirements Storage/Shipping Requirements
Temperature:            5 to 40oC  Temperature:            -20 to 70oC 
Relative Humidity:     less than  80% Relative Humidity:        less than 93%
The atmosphere pressure: 70 to 106 kPa The atmosphere pressure: 70 to 106 kPa

Electrical Specifications
Input Voltage Requirements
 AC Voltage:   110V~240V
 DC Voltage:   12V
 Inlet Battery:12V , 2Ah (Can be used for 8 hours)
 Frequency:     50Hz/60Hz
 Power:             2A

Pneumatic Specifications
Pneumatic Specifications 
Oxygen Inlet
Pressurized Oxygen:   0.4MPa medical oxygen or use 2L oxygen cylinder (Choice of suppliers)

Performance Specification And Tolerances
 Controls     
 Power    ON/OFF
 Modes   CMV(IPPV)   A/C(SIPPV)   SIMV   SPONT  PEEP(Choice of suppliers)
 Tidal Volume     50~1200ml
 Breath Rate  5~60 bpm 
 PEEP    0~2KPa 
 O2%      45~90% 
Sigh     1~10 per 100 breath
Peep   10-60 hpa  (Choice of suppliers)
Trigger Level    -10~0hpa
Over Pressure Relief    <6kpa 

Alarms and Indicators            
 High Pressure Alarm Limit  2~6kpa    flash, red indicator
 Low Pressure Alarm Limit    0.  2~2kpa    Red indicator 
Alarm Silence   120 sec
   Monitors            

Tidal Volume  50~1200ml
 Total Breath Rate     5~60 bpm 
Main Circuit Pressure -1~6KPa
Wave Display   Main Circuit Pressure
  What Is Provided            

 Part Name   Number   Specification 
  Ventilator   1    
 Oxygen Cylinder  1 2.5L can be used for 1 hour
 Breath Circuit  1 1.2m
 Carrying Package  1 all parts in it
 Breath Value 