Model Number |
721 |
721 Spectrophotometer Application: Apply to laboratory of Factory, Mine, Environmental Protection, Hospital, Epidemic Prevention and Scientific Research Department, quantificationally analyze the material content in the visible spectrum region.
Features: 1. Adopt Plane holographic grating as dispersive element, with high wavelength accuracy, T\A digital direct display. 2. Adopt advanced electronic circuit and elements, with high stability and sensitivity.
Main Technical Specification: 1. Wavelength Range: 330-1000nm 2. Wavelength Accuracy: +2nm or -2nm 3. Wavelength Repeatability: ≤0.5nm 4. Spectral Bandwidth: 6nm 5. Transmittance Accuracy: +0.5% or -0.5% 6. Transmittance Repeatability: ≤0.3% 7. Auto Adjust "0"\"100" 8. Power Supply: 90-260V\50-60Hz
|